DFI INFINITY
STRAIN METER/CONTROLLER
USER’S GUIDE
www.cooperinstruments.com
PH: 540-349-4746 FAX: 540-347-4755
Vista de pagina 0
1 2 3 4 5 6 ... 80 81

Indice de contenidos

Pagina 1 - USER’S GUIDE

DFI INFINITY STRAIN METER/CONTROLLER USER’S GUIDE www.cooperinstruments.com PH: 540-349-4746  FAX: 540-

Pagina 2 - CONTENTS

4 S /MAX BUTTON During run mode, pressing this button displays the “HI RDG” (peak reading) value that has occurred up to the moment the ‘MAX’ button

Pagina 3

4.2 Rear Of The Meter The following is a brief description of each part of the rear of the meter. The label on the top of the mounting sleeve (not th

Pagina 4

Figure 4-3 Rear View Figure 4-4 shows the rear of the meter with the optional BCD output board and a serial communications output board installed.

Pagina 5

3. To install optional output board(s) (Section 5.2.3). 4. To mount the meter in a panel (Section 5.2.5). 5.2 Assembly/Disassembly Opening Opening

Pagina 6 - 1.0 INTRODUCTION

Figure 5-2. Board Assembly Removing/Installing Detail The meter is now disassembled to the point where you can check and configure jumpers and instal

Pagina 7 - 2.0 UNPACKING

Figure 5-3. Transformer Jumpers 5.2.3 Printed Circuit Board(s) Installation To install optional printed circuit board(s): 1. “Reveal the Main Board

Pagina 8 - 3.0 SAFETY CONSIDERATIONS

Figures 5-5 through 5-13 show the layout of the seven (7) printed circuit boards with respective jumper blocks, where applicable, used in the meter. F

Pagina 9 - 4.0 PARTS OF THE METER

Figure 5-8. Relay Option Board Figure 5-9. 4-Relay Option Board The Analog option board has been updated. The figure shown is the latest version.

Pagina 10

Figure 5-11. RS-232 Option Board The RS-232 option board has been updated. The figure shown is the latest version. Some older versions of the RS-2

Pagina 11 - 4.2 Rear Of The Meter

2. You have configured all jumpers – those on the main board as well as those on any optional boards. See Section 5.2.2 for main board jumper config

Pagina 12 - 5.0 SETUP

CONTENTS NOTES, WARNINGS AND CAUTIONS ...1 1.0 INTRODUCTION ...

Pagina 13

The meter display should light, and pass through “RESET 2” to run or display mode. If the meter flashes an overscale or overload message, press the ‘M

Pagina 14

Jumpers for 0 to 10 V range: (meter supplied excitation) Jumpers for 0 to 100 V range: (meter supplied excitation) BRIDGE – BIPOLAR Jumpers for –5

Pagina 15 - 5.2.4 How To Access Jumpers

1.6 mA meter excitation (S2P, S2V) NOTES: 1) Maximum excitation for “Ratiometric” measurement is 10 Vdc. 2) Remove the “S2T” jumper when you use

Pagina 16

Figure 6.3. dc-Powered Main Board Jumper Positions S3 and S4 S1 On dc-powered units, side power board S1A Installed For 1.25 to 12 Vdc excitation

Pagina 17 - installed

Figure 7-2. Current Input With Sensor Excitation Figure 7-3 Voltage Input Without Sensor Excitation Figure 7-4 3-Wire Voltage Input With Sensor

Pagina 18 - 5.2.5 Panel Mount Assembly

Figure 7-6 Bridge Input With External Sensor Excitation Figure 7-7 Potentiometer Connections with Internal Power Supply & Ratio Measurement

Pagina 19

Figure 7-9. AC Connector Wiring at P1 Connect your AC meter power as described above and as shown in Figure 7-9. CAUTION: As mentioned in Section 5.

Pagina 20 - 6.0 JUMPER POSITIONS

If you received your meter and it is has been configured for an input other than what you require, you must proceed with the following steps before re

Pagina 21

8.3.1 SETTING INPUT CONFIGURATION (IN CNF) PRESS THEN PRESS (TO CHANGE PRESS IF REQUIRED) UNTIL DISPLAY SHOWS COMMENTS ‘MENU’ “IN CNF” Press the ‘

Pagina 22

Record your “INPUT1” and “READ 1” numbers, and your “INPUT2” and “READ 2” numbers, record them below and proceed as follows: Example INPUT1= 000003

Pagina 23 - 7.2 Signal Input Connections

9.2.2 RDG.CNF (READING CONFIGURATION)...34 9.2.3 RDG SC (Reading Scale) AND

Pagina 24

decimal point location within the “F’s” ‘MENU’ “CNT BY” Press the ‘MENU’ button to store your decimal point selection and the meter will momentar

Pagina 25 - 7.3 AC Wiring Connection

‘RESET’ ‘RESET’ “RESET2” then The Measured Value Press the ‘RESET’ the button two times. display will momentarily show “RESET2” and then will display

Pagina 26 - 8.1 Introduction

PRESS THEN PRESS (TO CHANGE IF REQUIRED) UNTIL DISPLAY SHOWS COMMENTS ‘MENU’ “IN.SC.OF” Press the ‘MENU’ button until the display shows “IN.SC.O

Pagina 27

These calculated scale factors are the numbers that your meter would display if you connected a signal source. You must complete the procedure for p

Pagina 28

‘MAX’ ‘MIN’ “000000.” Using the ‘MAX’ button to change the value of the flashing digit and the ‘MIN’ button to scroll to the next digit to the righ

Pagina 29

enter “–00020” in Reading Offset (“RDG OF”) and if your zero input shows a negative “–20” you would enter a positive 000020 in Reading Offset (“RDG OF

Pagina 30

‘MENU’ “STORED” the “IN CNF” Press the ‘MENU’ then button to store your selection. The display will momentarily show “STORED” then “IN CNF”. ‘RESET’

Pagina 31

MENU BUTTON MAIN MENU MIN/MAX* BUTTON SUB MENU CONDITION LOCKOUT CONFIGURATION #1 “L1C.1=0” Setpoint 1 change unlocked. “L1C.1=1” Setpoint 1 change

Pagina 32 - 8.4.2 Scaling Your Meter

“L2C.8=0” FIL CNF (adaptive/fixed filtering and for which output(s)) can be chosen. “L2C.8=1” FIL CNF (adaptive/fixed filtering and for which output

Pagina 33

MENU BUTTON MAIN MENU MIN/MAX* BUTTON SUB MENU DESCRIPTION LOCKOUT CONFIGURATION #4 “L4C.1=0” BAUD (communication rate) can be chosen. “L4C.1=1”

Pagina 34

13.8 Bringing Out The BCD Overflow Line ... 60 13.9 3 Digit At A T

Pagina 35

The 2-data-point method allows the user to use two known points to convert from one scale to another. For example, to convert from degrees Fahrenheit

Pagina 36

For “RDG OF”, you may choose to enter a reference temperature offset here (e.g., “-100.00”) so that the display will read deviation of the input from

Pagina 37

“INP.1=1” 50 Hz Read Rate: “INP.2=0” Slow “INP.2=1” Fast Input Voltage: “INP.3=0” Unipolar “INP.3=1” Bipolar Transmitter Type: “INP.4=0

Pagina 38

000000. (“100000.”) Enter second desired value. * The ‘MIN’ button allows you to sequence through “INPUT 1”, “READ 1”, “INPUT 2”, and “READ 2” hea

Pagina 39 - 9.2.1 INPUT

“010” “020” “050” “100” *Press the ‘MIN’ button to show “001”, “002”, “005”, “010”, “020”, “050”, or “100”. Press the ‘MAX’ button to select

Pagina 40

9.2.9 FIL TI (Filter Time Constant) Filter time constant is used to determine the number of readings the meter will average before displaying an inpu

Pagina 41

“SPC.3=1” Filtered Setpoint 2 Active zone: “SPC.4=0” Above “SPC.4=1” Below Setpoint 2 open-collector or relay output Active ON or OFF: “SPC.5

Pagina 42

Alarm 2 open-collector or relay output Active ON or OFF: “ALC.5=0” On “ALC.5=1” Off Filtered/unfiltered reading compared with Alarm 2 (Setpoin

Pagina 43 - 9.2.7 CNT BY (Count By)

The ‘MENU’ button stores the selected values for each “ALF” condition changed and advances the meter to the next configuration (“AL RDG”). Every und

Pagina 44

MENU BUTTON MAIN MENU MIN/MAX/MENU * BUTTON SUB MENU DESCRIPTION ALARMS 1 & 2 DEADBAND “0000” “AL DB”: (“0020”) Hysteresis for (with system deci

Pagina 45

18.3 Potentiometer Input ...70 18.4 G

Pagina 46

The ‘MAX’ button allows you to select the “0”, “1”, “2”, “3”, “4” or “5” state for each “OUT” condition. The ‘MENU’ button stores the selected value

Pagina 47

Press the ‘MAX’ button to select one of the above. Press the ‘MENU’ button to store the changes and the meter advances to the next configuration (“SE

Pagina 48

Data format allows you to select: • whether to transmit Alarm 1 or 2 status character [DAT.1] • whether to transmit peak and valley status characte

Pagina 49

• multipoint or point-to-point mode [BUS.4] • (if in point-to-point mode) select whether to communicate continuously or on command [BUS.5] • wheth

Pagina 50 - 9.2.18 BAUD (Baud Rate)

MENU BUTTON MAIN MENU MIN/MAX/MENU * BUTTON SUB MENU DESCRIPTION SERIAL COUNT This specifies the number of readings between data transmissions: “S

Pagina 51 - 9.2.21 DAT FT (Data Format)

10.1 Features Overview 1. Four full-range levels with many menu programmable features. 2. Independent operation or ganged action (including guard-ba

Pagina 52 - 9.2.22 BUS FT (Bus Format)

2. Press the ‘MENU’ button to see “L1 CNF” and then press the ‘MIN’ button to view “L1C.1=0” if “SP 1” is unlocked. If “L1C.1=1”, change to equal “0”

Pagina 53 - 9.2.23 SERCNT (Serial Count)

Figure 10-4. High Deviation for both Active Above and Active Below Figure 10-5. Low Deviation for both Active Above and Active Below Figure 10-6. B

Pagina 54 - 10.0 SETPOINTS/ALARMS

Figure 10-7. AL CNF Hysteresis 2.“ALC.2=0” turns the Alarm 1 (Setpoint 3) open-collector output ON when Setpoint 3 is active. “ALC.2=1” turns it OFF.

Pagina 55 - 10.2 Unlocking The Features

3.“ALF.3=0” makes “SP 4” INDEPENDENT, with a level equal to the value inserted for “SP 4”. “ALF.3=1” assigns “SP 4” to “SP 2”, placing it ABOVE “SP 2

Pagina 56

NOTES, WARNINGS AND CAUTIONS Information that is especially important to note is identified by these labels: • NOTE • WARNING • CAUTION • IMPORTAN

Pagina 57

10.11 Entering Setpoint Levels (In Run Mode) When you have completed selecting the setpoint(s) (and other features), the last press of the ‘MENU’ but

Pagina 58

Your meter converts display readings into an independently-scaled and-offset isolated voltage and/or current analog output. Isolation is accomplished

Pagina 59

3. Store this value by pressing the ‘MENU’ button and then advance to “READ2”. Use the ‘MIN’ and ‘MAX’ buttons to enter a large display value, for ex

Pagina 60 - 12.0 ANALOG OUTPUT OPTION

12.6 Filter Configuration FIL.CNF Value To Be Transmitted On Analog Output 1. Press the ‘MENU’ button until the display shows “FIL.CNF”. 2. Press th

Pagina 61 - 12.2 Unlocking

Figure 13-2 shows the board connections and pin designators. The locations of the jumpers are also shown. Figure 13-2. BCD Option Board 13.2 BCD Ca

Pagina 62

13.5 Selecting The Source Of BCD Data: “OUT .CNF” If “L3C.7=0” has been selected to unlock the Output Configuration byte, set “OUT.3=1” to send data t

Pagina 63 - 13.0 BCD OPTION

Figure 13-3. Address Programming Chart for 4-line Address NOTE: “X” in chart indicates jumper that must be installed. EXAMPLE: For a positive true a

Pagina 64 - 13.3 Interconnect Board

1-ampere Form-C electro-mechanical relays for 4 Relay Option). Each relay can accommodate a single setpoint. 200 W, 2500pf snubbers are provided for

Pagina 65

B,D A, C Assigns SP2 to Relay 2 (P7) Assigns SP3 to Relay 1 (P6) Assigns SP4 to Relay 4 (P18) B,D B,D Assigns SP1 to Relay 3 (P18) Assigns SP2 to

Pagina 66 - 14. RELAY OPTIONS

5 6 Common 4 NC4 (Normally Closed) 15.0 RS-232 OR RS-485 OPTION BOARD 15.1 Features Overview The Isolated Serial RS-232 Communications Board provide

Pagina 67

INF-1 ANALOG OUTPUT CARD W/ 4-20 MA/0-10V OUTPUT INF-2 RS-232 CARD INF-3 DUAL RELAY CARD INF-4 BCD ISOLATED PARALLEL OUTPUT INF-5 10-32 Vdc POWER

Pagina 68 - 14.2 Wiring/Connections

Figure 15-2a. Older RS-232 Option Board Figure 15-2b. Older RS-485 Option Board and Pin Designations and Pin Designations 15.2 Fr

Pagina 69 - 15.1 Features Overview

16.1 Tare (PIN 1) Tare is available when pin P2-1 and P2-4 are connected to a momentary contact switch. This feature allows you to automatically zero

Pagina 70 - 16.0 EXTERNAL CONTROL LINES

16.11 Print Command And/Or Reset Of Alarms (PIN 11) Grounding this pin to P2-7 when “BUS.8=1” has been programmed will initiate a meter printout via

Pagina 71

17.1 Error Mode Messages 17.1.1 Flashing 999999 (Numerical Overflow) The maximum number of counts in the display cannot exceed –99999 or 999999. If, b

Pagina 72

17.2 Troubleshooting Guide 99999 POSSIBLE CAUSE: Active decimal point change driving the display into numerical overload. TO CORRECT: Press the ‘MAX’

Pagina 73

TO CORRECT: Reduce the “CNT BY” count by value to 001. If you have an active decimal point selected, move the decimal point one or more positions to t

Pagina 74 - 17.2 Troubleshooting Guide

SPAN TEMPCO: less than 20 ppm/°C STEP RESPONSE: 1 second to 99.9% WARMUP: 55 minutes to rated accuracy It is recommended that the unit be continuou

Pagina 75 - 18.0 SPECIFICATIONS

0.001% of span/volt power supply rejection EXTERNAL FUSE REQUIRED: 115 Vac IEC 127-2/III,125 mA,250V (Time-Leg) or UL Slow Blow, 125mA, 250V 2

Pagina 76

DURING SETPOINT ADJUST (RUN MODE) OUT OF SELECTED DIGIT RANGE: “999999” NOT STORED IN EEPROM: “NOSTOR” VALUE PUT IN EEPROM: “STORED” 19.0 FAC

Pagina 77

Setpoints 1 & 2 Deadband “SP DB”: 0020 Alarms 1 & 2 Deadband “AL DB”: 0020 OUT.4=0 OUT.6=0 Output Scale and Offset “OT.SC.OF”: Bus Format “B

Pagina 78 - 19.0 FACTORY PRESET VALUES

1 Rear Protective Cover with Screw 1 20-Socket Ribbon Connector (P2 Connector) 2 Panel-Mounting Gaskets (1 Spare) 1 Strain Gauge Owner’s Guide

Pagina 79

For Pot: __________ Reading Configuration “RDG.CNF”: RDG.1=__ RDG.2=__ RDG.3=__ RDG.4=__ Reading Scale and Reading Offset “RDG SC” and “RDG OF” Readi

Pagina 80

CF 67 76 M1291/N/0403 11279ML-02 Rev. A Address “ADDRES” (for RS-485): ______ Data Format “DAT FT”: DAT.1=__ DAT.2=__ DAT.3=__ DAT.4=__ DAT.5=__ DA

Pagina 81 - WARRANTY REPAIR POLICY

• Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating. Use electrical wires with a