CF 195 Users Guide Model 7i Digital Force/Torque Indicator www.cooperinstruments.com PH: 540-349-4746 FAX: 540-347-4755
Model 7i Digital Force / Torque Indicator User’s Guide 9MODE Toggles between measurement modes. (DOWN) Navigates down through the menu a
Model 7i Digital Force / Torque Indicator User’s Guide 10Current Reading – Applies to the peak capture rate of the instrument. Displayed R
Model 7i Digital Force / Torque Indicator User’s Guide 11 – the displayed value is greater than the upper set point (NO GO HIGH) – th
Model 7i Digital Force / Torque Indicator User’s Guide 12To enable Break Detection and configure the automatic functions, select Break Dete
Model 7i Digital Force / Torque Indicator User’s Guide 13 7.3 Auto Output Settings Scroll to Auto Settings in the Break Detection menu and
Model 7i Digital Force / Torque Indicator User’s Guide 148.1 Real Time (RT) The primary reading corresponds to the live measured reading.
Model 7i Digital Force / Torque Indicator User’s Guide 15Parameter Description Initial Delay The time delay, in seconds, before the averagi
Model 7i Digital Force / Torque Indicator User’s Guide 169.5 External Trigger (ET) This mode of operation is useful for measuring electrica
Model 7i Digital Force / Torque Indicator User’s Guide 179.5.1 External Trigger Schematic Diagram 14GND1IN22kSWITCH UNDER TESTGAUGE+3.3VTO
Model 7i Digital Force / Torque Indicator User’s Guide 18 Select “+ More” for additional options: 9.6.3 Stop Condition Data capture i
Model 7i Digital Force / Torque Indicator User’s Guide 1Thank you… Thank you for purchasing a Mark-10 Model 7i force/torque indicator, de
Model 7i Digital Force / Torque Indicator User’s Guide 19Note 2: Data capture may be manually stopped at any time by pressing the ZERO key.
Model 7i Digital Force / Torque Indicator User’s Guide 20 9.7.2 Peak Settings The display appears as follows: Thresholds and percentage
Model 7i Digital Force / Torque Indicator User’s Guide 219.7.3 Auto Output Settings Scroll to Auto Output Settings and press ENTER to selec
Model 7i Digital Force / Torque Indicator User’s Guide 22the desired reading and press DELETE. The letter “D” appears to the left of the re
Model 7i Digital Force / Torque Indicator User’s Guide 23 If no option is selected, this screen will be displayed indefinitely, or until b
Model 7i Digital Force / Torque Indicator User’s Guide 2412 COEFFICIENT OF FRICTION The 7i, with an appropriate force sensor, can dete
Model 7i Digital Force / Torque Indicator User’s Guide 25 14 COMMUNICATIONS AND OUTPUTS Communication with the 7i indicator is achieved
Model 7i Digital Force / Torque Indicator User’s Guide 26 Selection Description Numeric + Units Output format includes the value and unit
Model 7i Digital Force / Torque Indicator User’s Guide 27 14.3 Analog Output This output can be used for chart recorders, oscilloscopes, d
Model 7i Digital Force / Torque Indicator User’s Guide 2814.5 I/O Connector Pin Diagram (DB-9HD-15 female) * Maximum vo
Model 7i Digital Force / Torque Indicator User’s Guide 21 OVERVIEW 1.1 List of included items Qty. Part No. Description 1 12-1049 Ca
Model 7i Digital Force / Torque Indicator User’s Guide 2914.6 Command Set / Gauge Control Language 2 (GCL2) The 7i may be controlled by an
Model 7i Digital Force / Torque Indicator User’s Guide 30 Memory & Statistics MEM Transmit all stored readings STA Transmit sta
Model 7i Digital Force / Torque Indicator User’s Guide 31 Other Commands AOFFn Auto-shutoff. n=0-30 minutes. 0=auto shutoff disabled
Model 7i Digital Force / Torque Indicator User’s Guide 32 Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/
Model 7i Digital Force / Torque Indicator User’s Guide 33 Optionally exercise the sensor several times (at full scale, if possibl
Model 7i Digital Force / Torque Indicator User’s Guide 34 Press ENTER. 11. At the completion of the tension calibration, the dis
Model 7i Digital Force / Torque Indicator User’s Guide 35Causes: 1. The calibration weight does not match the set value. 2. If using a
Model 7i Digital Force / Torque Indicator User’s Guide 3616.3 Locking Out Other Keys Other keys may be locked out individually. Select any
Model 7i Digital Force / Torque Indicator User’s Guide 37Selection Description Disabled Disable automatic shutoff. Enabled Enable automatic
Model 7i Digital Force / Torque Indicator User’s Guide 38 17.5 Initial settings This section is used to configure the initial settings up
Model 7i Digital Force / Torque Indicator User’s Guide 3Example 1 Model MR01-100 sensor with Model 7i Indicator MR01-100 ±0.15% of full sc
Model 7i Digital Force / Torque Indicator User’s Guide 3918 SPECIFICATIONS 18.1 General Accuracy: ±0.1% of full scale + sensorSampli
Model 7i Digital Force / Torque Indicator User’s Guide 40 Data Format Numeric + Units Auto Output Disabled Outputs per Sec
WARRANTY REPAIR POLICY Limited Warranty On Products Any Cooper Instruments product which, under normal operating conditions, proves defective in ma
Model 7i Digital Force / Torque Indicator User’s Guide 48. Typical materials able to be tested include many manufactured items, such as sp
Model 7i Digital Force / Torque Indicator User’s Guide 5off), a message appears, “BATTERY VOLTAGE TOO LOW. POWERING OFF”. An audio tone wil
Model 7i Digital Force / Torque Indicator User’s Guide 6on the printed circuit boards. 3.3 Mounting to a plate The 7i
Model 7i Digital Force / Torque Indicator User’s Guide 74 HOME SCREEN AND CONTROLS 4.1 Home Screen No. Name Description 1 Mea
Model 7i Digital Force / Torque Indicator User’s Guide 8 N – Newton kN – Kilonewton mN – Millinewton Torque units:
Comentarios a estos manuales