CF 112 1 Version 3.1-0
DFI 200 HAND HELD INDICATOR OPERATION MANUAL
1 INTRODUCTION
This manual contains information for the operation of the DFI 200 hand held load cell monitor. The DFI 200 is
suitable for measuring weight, pressure, force, torque etc. The input required by the DFI 200 is a 4 wire mV/V
output type transducer. 5V excitation is provided for the transducer. Calibration, and set up functions are easily
achieved via front pushbuttons.
The DFI 200 is a high accuracy meter with the capability of high speed sampling. Sample rate is programmable in
steps from 5 to 100 samples per second. The instrument has a full scale range of 0.5mV/V to 100mV/V. Up to 3
separate calibration sets can be stored internally allowing viewing of scaling for up to 3 different load cells, or 3
scaling ranges without the need to recalibrate each time.
When used with the external memory option the DFI 200 can “read” the calibration scaling values stored in memory
in the load cell connector shell. The instrument can then display the correct scaling value for each individual load
cell which uses a memory device without the need for recalibration.
The DFI 200 is supplied with RS232 communications as standard but by default this is disabled. See “Serial
communications” chapter for internal DIP switch settings required for enabling the serial communications.
The DFI 200 has three calibration/scaling method options. It may be calibrated by applying two known weights to
the load cell or by entering the mV/V value for the load cell or via a single offset value.
Vista de pagina 0
1 2 3 4 5 6 ... 20 21

Indice de contenidos

Pagina 1 - 1 INTRODUCTION

CF 112 1 Version 3.1-0 DFI 200 HAND HELD INDICATOR OPERATION MANUAL 1 INTRODUCTION This manual contains information for the operation of the D

Pagina 2 - 2 ELECTRICAL CONNECTIONS

CF 112 10 Version 3.1-0 Use of and as reference values. Note: do not use this facility if an optional external memory chip is connected. If using

Pagina 3 - CF 112 3 Version 3.1-0

CF 112 11 Version 3.1-0 Note that the zero operation is cumulative i.e. the zero operations are added in memory. For example if the display is scale

Pagina 4 - 4 EXPLANATION OF FUNCTIONS

CF 112 12 Version 3.1-0 Default Value: Power off time - selects the automatic power off time in minutes. This function allows the instrument to c

Pagina 5 - CF 112 5 Version 3.1-0

CF 112 13 Version 3.1-0 function characters (such as <STX> and <CR>). Therefore 32 (DEC) or 20 (HEX) is address 0, 42 (DEC) or 2A (HEX)

Pagina 6 - CF 112 6 Version 3.1-0

CF 112 14 Version 3.1-0 Note: each external memory device should be allocated a unit address, see function. Valid addresses are 1 to 250. If the

Pagina 7 - CF 112 7 Version 3.1-0

CF 112 15 Version 3.1-0 6 SERIAL COMMUNICATIONS 6.1 DIP switch settings The RS232 communications can be enabled or disabled via a DIP switch SW7 loc

Pagina 8 - CF 112 8 Version 3.1-0

CF 112 16 Version 3.1-0 6.3 Serial communications function settings and commands With the communications enabled the following extra functions will

Pagina 9 - CF 112 9 Version 3.1-0

CF 112 17 Version 3.1-0 (set RS232/485 interface mode) Select or . Allows user to select the RS232/485 interface operation as follows:- • - Send

Pagina 10 - 4.13 Zero range limit

CF 112 18 Version 3.1-0 A typical format for the host command is:- <STX>CA<CR> (Standard read) Where: <STX> is Start of Text Cha

Pagina 11 - CF 112 11 Version 3.1-0

CF 112 19 Version 3.1-0 4. Invalid Command. If the command received from the host is not valid then the unit may return the following:- <ACK>?

Pagina 12 - CF 112 12 Version 3.1-0

CF 112 2 Version 3.1-0 2 ELECTRICAL CONNECTIONS Power for the DFI 200 is provided by a 9VDC (216) battery. The battery cover is located at the unde

Pagina 13 - 5 EXTERNAL MEMORY

CF 112 20 Version 3.1-0 Note: Figures in the table above apply when the digital filter setting is 0. Add 0.5 bits effective resolution for each step

Pagina 14 - CF 112 14 Version 3.1-0

CF 112 21 Version 3.1-0 found and returned as is or merely recalibrated. It may be possible for out-of-warranty products to be returned to the exac

Pagina 15 - 6 SERIAL COMMUNICATIONS

CF 112 3 Version 3.1-0 3 FUNCTION TABLES – SUMMARY OF SETUP FUNCTIONS Functions in this first table are available in FUNC or CAL mode. Display Fun

Pagina 16 - CF 112 16 Version 3.1-0

CF 112 4 Version 3.1-0 4 EXPLANATION OF FUNCTIONS The DFI 200 setup and calibration functions are configured through a push button sequence. Two

Pagina 17 - CF 112 17 Version 3.1-0

CF 112 5 Version 3.1-0 scaling values may be entered for a single cell, allowing the same cell to display in different units e.g. Kilonewtons, Kilog

Pagina 18 - CF 112 18 Version 3.1-0

CF 112 6 Version 3.1-0 Default Value: Displays and sets the decimal point place. By pressing the or pushbuttons the decimal point position may

Pagina 19 - 7 SPECIFICATIONS

CF 112 7 Version 3.1-0 measurement mode. Any changes to the function selected when the button was pressed will not be saved. This function may be se

Pagina 20 - 8 WARRANTY

CF 112 8 Version 3.1-0 9. The display will now indicate (2nd calibration point). If you do not wish to enter the second point at this stage then p

Pagina 21 - CF 112 21 Version 3.1-0

CF 112 9 Version 3.1-0 4.11 Alternative mV/V calibration method Display: Range: Default Value: This function provides an alternative calibra