DFI INFINITY CS (V- M3598/N/0906) 18 CF 70
4.5 Selecting a Display Color
Refer to Table 6-1 for a summary list of menu configuration.
Selecting Display Color is not active unless your meter is a Version “B”.
To select a display color, follow these steps:
1. Press the MENU button until the meter shows
.
2. Press the X/TARE button. The meter shows one of the following:
3. Press the S/NT/GRS button to scroll between available choices.
Press the MENU button to store your choice. The meter momentarily shows
, followed by the next menu
(Setpoint 1 Configuration). Or you can press the RESET button to abort and go back to the menu.
4.6 Using Setpoint 1 Configuration
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Configuration
is not active unless your meter has dual relay output capabilities. The LED’s will
display whether the
is active or not. You may use Setpoint 1 Configuration for the following:
To set the setpoint’s active band above or below your chosen value
To select whether the setpoint operation is latched or unlatched
Assigning setpoint values to the net or gross reading
4.6.1 Setting Setpoint 1’s Active Band
1. Press the MENU button until the meter shows .
2. Press the X/TARE button. The meter shows one of the following:
= (Active above the setpoint) (Default)
= Active below the setpoint
3. Press the S/NT/GRS button to toggle between available choices.
4. Press the X/TARE button to select if Setpoint 1 is latched/unlatched or press the MENU button to store your
selection. The unit shows
.
4.6.2 Selecting if Setpoint 1 is Latched of Unlatched
1. Press the MENU button until displays, then press the X/TARE button twice.
or
Press the X/TARE button from
.
The meter shows one of the following:
= Setpoint 1 to be unlatched (Default)
= Setpoint 1 to be latched
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the X/TARE button to assign Setpoint 1 values to net or gross reading or press the MENU button to
enter
(Setpoint 2 Configuration).
4.6.3 Assigning Setpoint 1 Values to Net or Gross Readings
1. Press the MENU button until displays, then press the X/TARE button twice.
or
Press the X/TARE button from
.
The meter shows one of the following:
Vista de pagina 20