CF 183 18 32-1113 REV 2 1010
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will return to the
menu without saving changes.
4. After the number of calibration points has been entered, press ENTER. The display will appear as follows:
5. Place the force gauge horizontally on a level surface free from vibration, then press ZERO. The gauge will
calculate offsets, and the display will appear as follows:
5
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply any calibration
loads. Then press ENTER.
7. The display will appear as follows:
Vista de pagina 20