CF 155 6 9/25/07
5.1 STANDARD INSTRUMENTS
MODEL RANGE RESOLUTION STANDARD CABLE SELECTIONS
ACX-250 20-250 LBS 0.5 LB 1/16", 3/32", 1/8", 5/32"
ACX-500 40-500 LBS 1.0 LB 3/32" 1/8", 5/32"
ACX-CWT 20-250 LBS 0.5 LB CWT-5, CWT-10, CWT-15, CWT-20, CWT-25
ACX-CMB 20-250 LBS 0.5 LB 1/16", 3/32", 1/8", 5/32", CWT-5 thru CWT-25
6.0 AVAILABLE OPTIONS FOR ALL INSTRUMENTS:
Note: Add option designator to the end of the part number. (For example, a Model ACX-250-E denotes a standard,
ACX-250 with the optional RS-232, serial output function installed).
-A Analog Output option.
0-5 VDC or 4-20 mA with Software-definable ending sequences. Provided with
10’ sequences. Provided with 10’ cable. Data outputted at 40 Hz.
-E RS-233 Serial Output option.
Provided with 10’ cable to Interface with your receiving device. Select data
sampling rate from 1, 2 or 5 Hz.
Input Real-Time Serial Data into any Windows® application using the optional
software WinWedge Pro® for Windows, or configure your Window’s® operating
system to capture the data via Hyper Terminal®.
7.0 SPECIFICATIONS
Approximate weight is 1-1/2 lbs (Weight varies depending upon instrument configuration).
Re-chargeable NIMH battery with power supply provided. Approximately 6 hours of operation per battery charge
depending upon backlighting intensity.
Power Supply operates with input voltages from 100 – 240V and includes several, interchangeable “blades”
allowing use with European, US, Australian, and other plug configurations.
Automatic shutoff after 10 minutes of non-use.
Instruments may be operated continuously while connected to power supply.
Durable, lightweight carrying case with protective foam inserts.
CE Certification complying with heavy, industrial immunity standards.
8.0 SETUP
CHECKING ACCURACY. If you want to verify the accuracy of your instrument, simulate a tension load on your
cable by suspending known weights to a sample length of the same type of cable. When you engage the meter this
weight value should display. Thus when a 100 lb weight is freely suspended from the cable, the tension will be 100
lbs, and your instrument should indicate 100 lbs.
8.1 Custom Calibrations for the ACX-Series of Digital Tension Meters
Warning: The Custom Calibration feature is password protected. In order to perform a Custom Calibration a
series of key inputs are required. If you enter the password sequence and the press Enter while an existing
calibration in the menu is selected/highlighted, it will void that current calibration selection.
8.2 Performing a 5-Point, Multilinerization Calibration
Step 1: From the Main Display press or arrows until you reach Setup. Press ENTER. Press the arrow button
until you reach Cal Tension. Press ENTER. The screen will display: Setup, Calibrate Tension, Enter Password.